Why Do I Get Over 2 Foot-Candle / 20 Lux Visible Light from My LED UV-A Lamp?

  •  

A quick history of the visible light emissions requirement, where it is today, and an overview of UV meters  

By David Geis, Product Manager 

In both the Penetrant and Magnetic Particle checklists, Nadcap requires UV-A lamps to be checked for visible emission at 15 in / 38 cm using a standard light meter. Frequently, readings are above 2 foot-candles / 20 lux, and the conclusion is there’s something wrong with the lamp. However, the process of identifying the root cause is not that simple. 

 

Visible Light Emission Requirement 

The requirement for checking visible light emission from UV-A lamps traces back 25-30 years. At the time, there was only one type of inspection lamp – mercury vapor – but there were a couple of different configurations. The standard 100W lamp required a separate “woods glass” filter to block out visible light and only transmit UV-A. By far, this was the most common lamp in use, but it was not the only one available. 125W self-filtered bulbs offered high intensity and the bulb itself was “woods glass” to filter out visible light. The problem was the bulb glass was too thin to block all the visible light.  

 

The New Requirement 

In response, aerospace primes created a new requirement: if you were using a 125W lamp, you had to check the visible light emission to maintain the dark inspection area below 2 foot-candles / 20 lux. A simple solution, but there was some complexity involved – different light meters would give widely different results. That’s because the reading you get from a standard light meter depends on the filters used to tailor its response curve, and different manufacturers use different filter stacks. Typically that variation isn’t a problem when measuring a white light source, but not so with UV-A sources. Inspection lamps are very intense, and the filter stack in the light meter must block all UV energy and only pass the tiny amount of visible light remaining. Optical filters are not 100% effective, so the meter reading depends more on the exact types and tolerances of the filter stack than it does on the source itself. To solve this technical difficulty, aerospace primes included a table of approved light meters and the maximum acceptable reading for each one. The acceptable readings varied widely, from 0.3 to 8 FC or 3 to 86 lx depending on the light meter. 

In time, this aerospace prime requirement was included in Nadcap checklists, but there were a few modifications: 

  1. The requirement applied to all UV-A lamps, not just 125W self-filtered lamps. That means that a new quality check was required for a problem that did not exist in all lamps. 
  2. The test procedure was changed from a standard 1,200 µW/cm2 lamp intensity to a standard 15 in / 38 cm distance. Originally, the lamp was held at whatever distance was needed to get 1,200 µW/cm2, and then the visible light was checked. Now the lamp is at a fixed distance, meaning higher intensity lamps produce higher readings regardless of the relative light emission. 
  3. The table of light meters was not included. All light meters were assumed to have the same response curve, even though that was not true. 

Standard 100W mercury vapor lamps produced a very sharp peak at 365 nm, so there were seldom any problems with visible light emission. But LED lamps changed things. 

UV-A LEDs have a broader emission spectrum, including a small amount of visible light, which can be observed as a deep violet glare. To eliminate that glare, ASTM E3022 requires that inspection lamps include a UV-A Pass filter to block visible emission. Learn about the function and importance of UV-A pass filters, and why ASTM E3022 requires all LED UV lamps to have them here. Even so, many operators still get readings above 2 FC / 20 lx with their standard light meters, and that continues to be a problem. 

Conclusion 

Ultimately, the answer is that standard light meters are not able to accurately measure the small visible light emissions from a high-intensity UV-A source. LED lamp manufacturers use spectroradiometers to calculate the visible emissions from each lamp during ASTM E3022 certification. With standard light meters, the filter stacks and tolerances vary from model to model and from manufacturer to manufacturer. That means different light meters give you different readings from the same LED lamp, dependent on each meter’s response curve. While it didn’t make much difference with old mercury vapor lamps, it does make a difference with the broader emission spectrum of UV-A LEDs. 

Please wait while we gather your results.

Related Blog Posts

What You Need to Start-up In-house Testing [Checklist]

What You Need to Start-up In-house Testing [Checklist]

Step-by-step process listing everything from the equipment needed in magnetic particle testing and liquid penetrant testing to preparation for NDT audits

Read More...

Learn how to achieve a high intensity beam of light with UV LEDs for NDT inspection

NRTL: What Does It Mean and Why Does It Matter?

NRTL stands for Nationally Recognized Testing Laboratory, but what does that mean to you?

Read More...

Shedding Light on ASTM E2297 & E3022: A Guide to NDT UV Lamp Standards

An overview of two of the most common standards for NDT UV light sources, ASTM E2297 and ASTM E3022.

Read More...

How to Select the Right UV-A Lamp for Your NDT Needs [Checklist]

4 things to consider when looking for a new LED UV light for fluorescent penetrant inspection or magnetic particle testing.

Read More...

Understanding Safety Data Sheets

Understanding Safety Data Sheets: Do You Know What’s in the Chemicals You Handle?

We highlight the 5 major sections to understand on a Safety Data Sheets (SDS) in accordance with US OHSHA Hazcom 2012 GHS regulations

Read More...

Understanding the Magnaflux Shelf Life Statement for NDT Chemicals

Understanding the Magnaflux Shelf Life Statement for NDT Chemicals

Answering popular questions regarding shelf life for magnetic particle and liquid penetrant products used in nondestructive testing

Read More...

Why Does ASTM E3022 Require a Filter for LED UV Lamps?

Learn about the function and importance of UV-a pass filters, and why ASTM E3022 requires all LED UV-A lamps to have them.

Read More...

Handheld LED UV Lamps

About the EV6000 Handheld LED UV Lamp

Check out how Magnaflux’s EV6000 stacks against other leading UV-A portable handheld lamps in the market

Read More...

EV6000 Hand-held UV Lamp Demo and Q&A [Webinar]

Watch and listen to a 20-minute EV6000 demonstration by two outstanding NDT experts who afterwards host a 20-minute Q&A for participants to ask their own LED UV questions.

Read More...

ST700 Hood Mount is a freely-adjustable mounting frame

New Mounting System for Overhead UV-A Lamps

The ST700 Hood Mount is a freely-adjustable mounting frame to give operators the option to position their overhead UV-A Lamp to optimize inspection ease

Read More...

Magnaflux ST700 Overhead NDT Inspection Lamp

Stationary LED UV Lamp for NDT Inspections

See how the ST700 UV inspection lamp​ can eliminate the most common UV illumination challenges fluorescent NDT testing

Read More...

Learn how to achieve a high intensity beam of light with UV LEDs for NDT inspection

3 Facts About LED Optics and The Impact of Ultraviolet Light

Learn how to achieve a high intensity beam of light with UV LEDs for NDT inspection

Read More...

4 Monthly Checks Required for all LED Inspection-Grade Lamps

4 Monthly Checks Required for all LED Inspection-Grade Lamps

User-check requirements for LED UV-A lamps, and how to plot the beam of an LED UV-A lamp for Airbus AITM6-1001 Issue 11

Read More...

Complying with Airbus AITM requirements – How to check UV stability

Complying with Airbus AITM Requirements – How to Check UV Stability

Learn how to comply with the new requirements of the latest revision of Airbus AITM 6-1001

Read More...

Defining IP Ratings for Electronics Used in Mag Particle Testing & Liquid Penetrant Testing

Defining IP Ratings for Electronics Used in Mag Particle Testing & Liquid Penetrant Testing

In this article, we break down the ANSI/IEC 60529 Ingress Protection Code with a chart to show what's essential for NDT equipment

Read More...

NDT UV LED Lamps

Frequently Asked Questions About NDT UV LED Lamps

Frequently Asked Questions About NDT UV LED Lamps

Read More...

Nadcap AC7114 Rev J

How to Comply with NADCAP AC7114 Requirements for LED UV Lamps

An overview of Rev J changes to Nadcap AC7114 Non-Destructive Testing audit checklists for Penetrant and Magnetic Particle

Read More...

Airplane Engine NDT

How to Improve Fluorescent NDT Process Control with LED UV Lighting [Case Study]

See how an aerospace OEM manufacturer increased NDT inspection reliability & control with wide-beam, overhead LED UV-A lamps

Read More...

ST700 High Volume Inspections

How to Increase High-Volume NDT Inspection Speed and Reliability with LED UV Lighting [Case Study]

See how an automotive parts manufacturer decreased inspection time and improved reliability with wide-beam, overhead UV lamps

Read More...

Dye Penetrant Testing on Pipe

NDT UV Lamps “Stabilize” vs “Warm Up”, What’s the Difference?

Understanding the difference between NDT mercury-vapor lamp “warm-up” time and LED black light “stabilization” time

Read More...

New High-Intensity, Dual-Light LED UV Lamp for NDT Pros

Learn about how the EV6500 UV lamp for nondestructive testing improves visibility & versatility while minimizing time to inspect parts

Read More...

ST700 Overhead Inspection UV Lamp

New Stationary LED Inspection UV Lamp for NDT Pros

Learn how this new UV lamp for nondestructive testing solves the biggest challenges in fluorescent NDT UV illumination.

Read More...

UV-A Intensity: What’s Special About 5,000 µW?

UV-A Intensity: What’s Special About 5,000 µW?

Most UV-A lamps on the market are set to provide 5,000 µW/cm2 at a distance of 15 in / 38 cm. What’s so special about those numbers?

Read More...

Why Do I Get Over 2 Foot-Candle / 20 Lux Visible Light from My LED UV-A Lamp?

Why Do I Get Over 2 Foot-Candle / 20 Lux Visible Light from My LED UV-A Lamp?

A quick history of the visible light emissions requirement, where it is today, and an overview of UV meters

Read More...

JOIN OUR MAILING LIST TO RECEIVE THE LATEST NDT INSIGHTS AND ARTICLES FROM MAGNAFLUX

Magnaflux

155 Harlem Avenue
Glenview, IL 60025, USA
Telephone: +1 847-657-5300
Contact Magnaflux Customer Service

Select Your Country North America Mexico Brazil China Europe India New Zealand, Australia, Japan, Southeast Asia
© 2024 Magnaflux - All Rights Reserved.
top